Yes. Microedge provides several ways for third-party software, industrial systems, and custom applications to access data from Microedge data loggers.
The available method depends on the data logger model and your application requirements.
1. SiteView SDK
The SiteView SDK allows third-party applications and custom software to communicate with compatible Microedge data loggers programmatically.
Developers can use the SDK to integrate data logger communication, data download, and other supported functions into their own applications.
For detailed information, please visit:
SiteView SDK
https://www.microedgeinstruments.com/siteviewsdk.php
2. Modbus TCP with PRECISE-LOG Data Loggers
PRECISE-LOG data loggers can operate directly as Modbus TCP servers.
When Modbus TCP is enabled, compatible Modbus software or hardware can communicate with the logger over the network and read supported Input Registers and Holding Registers.
This allows PRECISE-LOG data loggers to be integrated with systems such as:
- PLCs
- HMIs
- SCADA systems
- Modbus software
- Other Modbus TCP master devices
For instructions, refer to:
How to Enable Modbus TCP on a PRECISE-LOG Data Logger
https://microedgeinstruments.com/meifaq/?p=680
3. Smart-Hub for USB-Based Data Loggers
For USB-based data loggers such as SITE-LOG and iLOG, Smart-Hub provides remote network access and standard interfaces for integration with other software and systems.
Connect the data logger to one of Smart-Hub’s USB ports and connect Smart-Hub to your Ethernet network.
Smart-Hub provides three main communication functions:
USB Device Server
Smart-Hub acts as a USB Device Server, allowing Microedge systems to remotely access data loggers connected to its USB ports.
Supported Microedge systems include:
- SiteView
- SiteView-Web
- SiteView SDK
- Pluto
This allows USB-based data loggers to be located remotely from the computer or server running the Microedge software.
RESTful Web API
Smart-Hub provides RESTful Web API endpoints for integration with third-party software and systems.
Through the API, an external application can access functions such as:
- View data logger activities
- Download logged data
- View real-time readings
This provides a standard web-based interface for integrating Microedge data loggers into other software platforms.
Modbus TCP Gateway
Smart-Hub also operates as a Modbus TCP Gateway.
Data loggers connected to Smart-Hub’s USB ports are presented as Modbus TCP slave devices, allowing Modbus TCP masters to access them over the network.
Real-time readings are stored in Input Registers and can be polled by:
- PLCs
- HMIs
- SCADA systems
- Modbus software
- Other Modbus TCP master devices
This allows USB-based Microedge data loggers to be integrated into Modbus systems without requiring the data loggers themselves to have built-in Ethernet or WiFi connectivity.
For more information, please visit:
Smart-Hub – 3-in-1 Communication Gateway
https://www.microedgeinstruments.com/accessories.php/smart-hub