If a data logger is configured to start logging at a future time, you can use the logger’s Activation button to manually start the logging session in the field.
1. Configure a Delayed Start
In SiteView, open the logger’s Configuration dialog.
Set Time to Start to a future date and time that will not normally be reached during your application.
Click OK to save the configuration to the logger.
The logger will now remain in Start Delay mode until it is activated.
2. Start Logging with the Activation Button
When you are ready to begin logging:
- Press and hold the Activation button on the logger.
- Continue holding the button until the status LED flashes once.
- Release the button.
The logger will then begin recording data immediately.
Important
Once the logger has already started logging, the Activation button no longer functions as a logging-start control.
During an active logging session, the button is used to reset the logger’s microcontroller, for example if the logger experiences a communication problem.